TandD MCR-4TC

  • 4 Channel Thermocouple Logger
  • Coupling Units Allows for Up to 16 Channels
  • Compatible with J,K,R,S & T types
  • -270° to +1760°C Measurement Range
  • 10Hz to 1 Hour Recording Interval Range
  • 960k Readings Internal Memory
  • SD Card Memory Expansion
  • Runs on 2 AA Alkaline Batteries

$367.00

In Stock

Description

The MCR-4TC Temperature Data Logger is a four channel battery operated temperature data logger designed to measure a wide range of temperatures using a thermocouple sensor. Up to 4 units can be coupled together making it possible to measure and record up to 16 channels.

Users can easily switch the LCD display between real-time measurement and the trend graph display.

Using a USB connection, recorded data can be sent directly to a PC, or a SD card (not supplied) can be used. Data can also be viewed and checked on the handy LCD display.

The MCR-4TC is convenient for recording temperatures in pipes and ducts, cooking equipment such as fryers and ovens, and recording boiler temperatures. Also great for management of refrigerated and frozen goods, or temperature testing of the electronic substrates.

Specifications

DimensionsH 120 mm x W 75 mm x D 32 mm

    • *1 : MCR-4TC has superior noise filter, but the measurement may sometimes fluctuate due to strong noise. Especially when the recording interval is set to 200 ms or less, the filtering becomes weaker and hence the fluctuation may become greater.
    • *2 : If the logging capacity is not filled at the end of one recording session, the logger can record up to 30 times.
    • *3 : Group Recording may not be started depending on the recording or measurement interval specifications of the connected Master unit.
    • *4 : Please check the Functions for information on memory cards whose operation has been confirmed.
    • *5 : Battery life varies depending upon the ambient temperature, the recording interval, the number of measurement channels, and the frequency of data export to a memory card.
MCR-4TC
Measurement Channels Temperature 4 ch
Compatible Sensors Thermocouple: Type K, J, T, S, R Measurement Units °C, °F Measurement Range Type K : -270 to 1370 °C
Type J : -210 to 1200 °C
Type T : -270 to 400 °C
Type S : -50 to 1760 °C
Type R : -50 to 1760 °C
Input Impedance Approx. 1 MΩ Accuracy (*1) Thermocouple Measurement
Type K, J, T: ±(0.5 °C + 0.3 % reading) [ at -100 °C or above ]
Type S, R : ±(1.5 °C + 0.3 % reading) [ at 100 °C or above ]
(Individual sensor inaccuracies not included.)
Cold Junction Compensation
±0.5 °C [ in operating environent of 10 to 40 °C ]
±0.8 °C [ in other operating environment ]Auto : According to the range in use
Measurement Resolution; 0.1 °C Recording Method Instantaneous or Average Recording Interval 100, 200, 500 ms.
1, 2, 5, 10, 15, 20, 30 sec.
1, 2, 5, 10, 15, 20, 30, 60 min.
Logging Capacity (*2) When recording 1 channel : up to 960,000 readings/ch
When recording 2 channels : up to 480,000 readings/ch
When recording 3 channels : up to 320,000 readings/ch
When recording 4 channels : up to 240,000 readings/ch
Recording Mode Endless ( Overwrite the most recent data when capacity is full )
One Time ( Stop recording when capacity is full )
Recording Start Method Immediate Start or Programmed Start ( by individual logger or by group ) Group Recording Up to 4 units ( 16 channels ) can be recorded simultaneously.
( Coupling of MCR-4TC and MCR-4V is possible. (*3) )
LCD Display Items Measurements, Trend Graph, Battery Level, etc. Communication Interfaces USB Communication Communication Time Download Times for Full Data via USB
– From the master or a single unit: Approx. 1 minute 30 seconds
– From a slave unit: Approx. 4 minutes 30 seconds
External Memory SD Memory Card, SDHC Memory Card (*4) ( For Manual or Automatic Data Export ) Power AA Alkaline Battery x 2 ( AA Ni-MH batteries may also be used ), USB Power ( 5 V 250 mA ) Battery Life (*5) For 4-Channel Measurement (AA Alkaline Battery)

Rec.Method
Instanteneous
Value
Average
Value
Rec
Interval
100 ms Approx. 5 days Approx. 5 days
500 ms Approx. 7 days Approx. 7 days
1 sec Approx. 21 days Approx. 7 days
5 sec or longer Approx. 60 days Approx. 21 days

(Battery life for 1-channel measurement is about 1.4 times longer than that of 4-channel measurement.)

Input Terminal Screwless Terminals
< Compatible Wires >
Single Wire : ø 0.32mm to ø 0.65mm ( AWG 28 – 22 )
Twisted Wire : 0.08mm2 to 0.32mm2 ( AWG 28 – 22 ), ø 0.12 mm or more in diameter
Stripping Length : 9 to 10 mm
Isolation CH1, CH2, CH3, CH4, and USB are isolated.
( Battery terminals are not isolated from the CH1-CH4 input terminals. )
CH1-CH4 Maximum Applied Voltage : ±50 V
Electrical Isolation Resistance : 50 MΩ or more ( DC ± 250 V )
Weight Approx. 190 g ( including batteries )
Operating Environment Temperature: 0 to 50 °C
Humidity: 90 %RH or less ( no condensation )
Accessories AA Alkaline Battery x 2, USB Communication Cable (US-15C), Software (CD-ROM), Card Slot Cover, User’s Manual Set (Warranty Included)